Background: Recent years, PET/CT equipment has played an increasingly important role in the medical field, and its quality control and evaluation requirements have become more stringent. Correspondingly, the performance testing phantom used for PET/CT quality control needs to be upgraded and optimized to meet the requirement of equipment imaging quality testing. The commonly used NEMA IEC Body performance testing phantom has the defects that it cannot detect micro-lesion and cannot measure CT performance parameters. This article proposes a NIM PET/CT phantom capable of simultaneously testing the performance of PET and CT equipment, and evaluates its imaging quality.
Methods: Compared with the NEMA IEC Body phantom, the PET performance testing module in the phantom has new balls with inner diameters of 4mm and 7mm. Combined with the CT performance testing module, it can be used for both PET and CT performance testing. The 28mm and 37mm balls are filled with pure water as a cold stove, and the remaining balls are filled with F-FDG solution as a hot stove. The activity concentration ratio of the hot ball to the background is 4:1. This study compares the imaging results of the NIM PET/CT phantom and the NEMA IEC Body phantom to verify its effectiveness; compares the imaging results of 3 different brands of PET/CT on the NIM phantom to verify that it is on different equipment The generalization ability of the system; apply PSF and TOF technology to the reconstruction algorithm and compare the improvement of the image quality; finally, the accuracy of the CT low-contrast module and the uniformity of the background are verified.
Results: The imaging quality of the NIM PET/CT phantom and the NEMA IEC Body phantom is relatively consistent. The NIM phantom under different types of PET/CT scans can detect 7mm balls without affecting the imaging quality of other areas, which is better The device can detect 4mm small balls, which can clearly classify the ability of different devices to present images of small lesions; the integration of PSF technology into the reconstruction algorithm significantly improves the image resolution and hot bulb contrast, but the edges of the lesions are still blurred , TOF technology improves the detection limit of the equipment and improves the overall quality of the image. PSF&TOF technology combines the advantages of the two, significantly reducing image noise and strengthening image details, so that the image quality has been comprehensively improved; the measurement in the CT module The result is consistent with the true value, and the relative error is within ±5%. The CT value transition between the background and pure water area is smooth, and the background uniformity is good.
Conclusion: NIM PET/CT Phantom and NEMA IEC Body phantom are comparability, and the former includes all the functions of the latter. In addition, the phantom can meet the testing requirement of different grades of PET/CT.