We determined backscatter factor (BSF) in a pediatric phantom using nanoDots OSLDs. NanoDots exhibit good batch homogeneity (< 5%) and reproducibility (3.3%), along with a linear dose response, but have been reported to have high angular dependence (up to 70%) in low kVp techniques such as mammography [25]. According to the study by Al-Senan [25] energy dependence was approximately 60% and recommendations were made for applying correction factors across the investigated energy range. Taking into consideration the angular and energy dependences and avoiding uncertainties associated with accumulated dose, OSLDs (nanoDots) are suitable for use as point dosimeters in diagnostic settings [25]. In the present study, we calibrated the nanoDots with respect to tube voltage to account for energy dependence (air kerma was calculated by multiplying the reading by a calibration constant for tube voltages of 50, 80 and 120 kV). Energy dependence was mitigated by calibrating each element with the energy used, but was possibly underestimated due to the nature of angular dependence. Nonetheless, we believe that underestimation may not be a significant issue for flat surfaces.
In the present study, increasing tube voltage and field size both increased the BSF values for the pediatric phantom, as shown in Table 2. The effect of tube voltage was more pronounced than that of field size, as the BSF consistently increased with tube voltage until reaching a saturated value. On the other hand, the BSF was converge to a saturation value when increasing field size as shown in Fig. 4. These findings align with existing BSF data [8], which demonstrated the highest BSF in lung tissue at 120 kVp, 3.0 mm Al, 0.1 mm Cu filter, and 6.62 mm Al HVL. This finding underscores the impact of HVL and filtering on BSF when tube voltages are fixed. However, differences emerged when these pediatric results were compared with adult BSF values [16]. In adults, the BSF increased until reaching a particular field size and then either remained unchanged or decreased with further increases in field size, which was presumably due to the 10% standard accuracy of nanoDots [22].
Discrepancies in BSF values can arise from various factors, and are not limited to tube voltage and field size. Beam quality also affects the BSF factor. Existing data [8], derived from a cuboidal mathematical phantom with three materials (water, ICRU, and PMMA), demonstrated greater BSF values than pediatric values across tube voltages and field sizes. This difference is attributed to variations in filtration and HVL, with increased filtration and HVL leading to higher BSF values. These findings corroborate the observations of Johns et al., who found that at a given HVL, kilovoltage and filtration influence the measured backscatter, with increasing HVL inducing higher BSF due to stronger beam qualities [26].
As shown in Table 1, the BSF is influenced by variations in phantom characteristics such as tissue density, thickness, and diameter, which is particularly relevant when considering phantoms of various compositions. For instance, the adult female RANDO phantom was designed with a focus on radiotherapy uses rather than being specifically tailored for diagnostic radiology. Consequently, its composition slightly differs from that of the human body and pediatric anthropomorphic phantoms, which were designed specifically for use in diagnostic radiology. The pediatric phantom includes materials that closely imitate human tissue in relation to factors like linear attenuation coefficient, physical characteristics, and electron density [24]. In the present study, we assumed that differences in diameter and thickness of the materials would have a more significant impact on the BSF in the pediatric and adult phantoms compared to the effects of their respective compositions. To validate this assumption, we measured the diameter and thickness of a pediatric Olivia phantom and an adult female RANDO phantom. The values for thickness were also less in the pediatric phantom: 15 cm vs. 18 cm for the skull, 12 cm vs. 19 cm for the chest, and 12 cm vs. 19.7 cm for the pelvis. Furthermore, with regard to accuracy, Benmakhlouf highlighted the necessity of performing material and thickness corrections of the irradiated objects when dealing with large field sizes [27]. These factors should be taken into consideration when interpreting the results of the present study and their implications for radiation dose measurements.
The shape and surface characteristics of a phantom also exert an influence on BSF. In adults, the breast region possesses intricate surface geometries that feature a valley-like configuration that is affected by transverse scattering of photons, thereby impacting the BSF [16]. However, the pediatric breast region has a smooth and flat surface. Accordingly, there is little to no effect of transverse scattering of photons in BSF calculated for a pediatric phantom in comparison to that for an adult phantom. Conversely, the pelvis does not have the valley-shaped expression of the chest, thus reducing the lateral scatter. However, when utilizing the nanoDots on the surface of the pelvis, there is a tendency for the dose to be underestimated due to the directional attributes of the dose reduction at 90° caused by transverse scattering, as reported by Al-Senan [25]. Whereas the prior BSF data [8] were derived for adults via the Monte Carlo method using a flat-surfaced cuboid phantom measuring 30 cm × 30 cm × 15 cm, we employed OSLDs and a direct measurement technique in the present study. Our study aimed to ascertain entrance air kerma at the surface and incident air kerma in the air at the same position without the pediatric anthropomorphic phantom present, and found that distinct BSF values emerged between the two groups. These findings underscore the importance of employing pediatric-specific BSF values when calculating ESD for pediatric cases.
The BSF values determined by Petoussi-Henss [8] may not align with BSF data derived from an anthropomorphic phantom, as evidenced in the comparison with BSF values acquired by Arimoto and Asada [16]. The present study undertook measurements and comparisons of BSF values between the pediatric anthropomorphic phantom (Olivia) and an adult phantom (RANDO), examining trends and correlations among factors including phantom type (pediatric vs. adult), tube voltage, and field size. BSF displayed an increase as tube voltage increased across all tested body regions, and exhibited increases for specific field sizes at certain tube voltages (Table 2, Fig. 3). In the pediatric phantom, BSF was lowest in the chest at 50 kV, indicating a minor backscattering effect in lung tissue compared to soft tissue, and peaked at 80 and 120 kV in the pelvis. Conversely, the adult phantom exhibited the highest BSF in the chest across all tube voltages due to its unique valley-like geometry, which is more susceptible to transverse scattering.
In addition, when comparing BSF values derived from the present pediatric data and existing data [8], the most differences were observed at 50 kV (11.0%), 80 kV (18.1%), and 120 kV (19.6%). These may arise from differences in HVL and other influence factors, which cause variation in beam quality and thus BSF values. It can be assumed that aluminum and copper filters enhance beam quality by compensating for low-energy components and reduce BSF. Investigating the influence of filters on BSF values is a subject of future inquiry. However, even though the differences have been found and showed that the BSF in children are slightly lower than in adults and as the difference between adult and pediatric values is in the range of some percent, so clinically, among the differences between two group the adult BSF is still possible for use to calculate ESD in pediatric.
There are some limitations of the present study. The results showed a difference in BSF values between the two groups in a limited scope of only three body regions, and thus the present findings cannot be applied to all regions of the body. Moreover, despite the presence of significant differences between the groups, it is challenging to draw firm conclusions about variations in trends because of the limited amount of data available. To accurately determine the trend of significant differences, it is necessary to conduct further research that includes other body regions, with an expanded amount of data.
However, BSFs have been calculated using a water phantom as the subject for more than 30 years. In DRLs for general radiography, ESD including a BSF is adopted as the DRL quantity. To estimate the ESD more accurately, it was necessary to use an anthropomorphic phantom instead of a water phantom, and to update the BSF, which was the original motivation for conducting the present experiment. In terms of the pediatric skull, we found that there was bone absorption and the BSF was considerably low, which are new findings. In addition, with the lack of available pediatric-specific BSFs, adult BSFs have been used to calculate pediatric ESD instead. We consider that the present finding of the importance of using a pediatric-specific BSF is of great clinical relevance.