Caustic pattern as fluorescence 3D image encoder
Figure 1a shows the illustration of the optical setup used for data acquisition in the proposed method. It is based on utilizing high spatial-frequency preserving, laterally (x-y)-invariant but depth (z)-variant caustics as the optical image encoder. The method preserves the high contrast optical signals from fluorescent sources located over a large depth range. For instance, Fig. 1b shows that the caustic signal remains above the noise floor for up to a defocus depth > 175 µm, while imaging with a 0.5 numerical aperture (NA) objective lens. While the preserved contrast presents an opportunity to perform extended depth-of-field microscopy, the caustic’s z-variant nature allows precise depth-encoding of these sources (in a single 2D wide-field image). The axial resolution and depth-localization accuracy of this computational microscopy method depends upon the NA of the objective lens. For instance, for a 0.5 NA objective, this caustic ring’s radius grows by 200 nm when the depth is shifted by 0.5 µm (Fig. 1c). The resulting accuracy with which the location can be resolved in z is 0.5 µm.
Microscope objectives are designed to minimize aberrations, and hence in the above experiments we employ a 1.1 mm thick parallel glass plate covering the sample to introduce spherical aberration and generate the caustic patterns in the defocus volume. When this glass plate is inserted between the sample and the objective lens, in an epifluorescence microscope, as shown in Fig. 1a, the rays incident at larger angles suffer larger displacements, as compared to rays incident at smaller angles and this phenomenon introduces the spherical aberration in the optical wave-field. The combination of spherical aberration and defocus results in a rotationally symmetric caustic pattern. The effect of the glass plate on the optical-field is negligible for low NA objective lenses and caustics becomes noticeable only above the NA around 0.40 (see Supplementary Figure S1).
When compared to the disc (Gaussian) defocus PSF observed in conventional imaging – without the glass plate - where there is minimal spherical aberration at the sample, CausWEI images exhibit a strong and clear retention of high spatial frequencies (see Supplementary Figure S2). Since the point spread function is a sharp circle, different depths can be digitally reconstructed by deconvolution with an analytically defined circular PSF of appropriate radii. Figure 1e-g and 2 shows the single-shot capture of objects located at various different defocus depths and their corresponding reconstructions.
Quantitative assessment of single-shot 3D reconstruction of fluorescent beads dispersed in PDMS
We first establish the depth-range and accuracy of depth-localisation of this imaging method with 10 µm fluorescent beads dispersed in ~ 1 mm thick PDMS (Polydimethylsiloxane). Figure 1e shows a single-shot wide-field 2D image with optical signal from beads located at various depths. From this measurement, a reconstruction of 3D volume with 100 depth slices was obtained, out of which 4 slices have been shown in Fig. 1g. Supplementary Video 1 shows the visualization of full 3D reconstructed volume. This result experimentally demonstrates that our method enables a scan-free imaging of objects located as deep as 175 µm defocus depth. To demonstrate the accuracy of depth-localization, we experimentally measure the change in the radius of the circle PSF with respect to a change in the depth of the source (calibration bead here). To achieve this, we estimate a value of the circle radius which produces the best focus in the reconstructed image. With a 0.5 µm change in depth, this estimated PSF grows by a radius of 200 nm. Both the depth-range and depth-localization values were specific for the 0.5 NA objective and will change with the NA of the objective. A lateral resolution of < 3 µm was experimentally measured, for an objective lens of 0.5 NA. Supplementary Figure S3 shows experimental data with 10 µm beads and ~ 3 µm nuclei. Confocal microscopy, to capture a 3D sample located across a 115 µm depth range in a clear background, would require scanning in both lateral and depth directions. To capture the same data volume at the same lateral resolution as CausWEI captured in one shot, a confocal microscope has to sequentially scan 7.3 x 106 voxels (lateral field-of-view x depth slices = 206 x 154 x 230 samples). Single-shot capture allows a data compression by a factor of the number of reconstructed depth slices, for example 230x in this example. Although our lateral resolution may be lower than the theoretical diffraction limit for a given objective lens, the depth localization accuracy is comparable to the theoretical axial resolution of confocal microscopy with the same NA objective.
To demonstrate the 3D imaging of extended objects, we consider an aggregate of fluorescent beads of size distribution 9.8 ± 0.5 µm in a PDMS medium. Figure 2a shows a caustic pattern image from this 3D object captured in one of the defocus planes. A reconstruction of 20 different slices was obtained and has been shown in Supplementary Video File 2. 5 out of these 20 slices showing the digital refocusing of different beads located at different depths are shown in Fig. 2b. The average diameter of the beads in this reconstructed aggregate is estimated to be 9.4 ± 0.5 µm, this includes measurements from 5 different depths.
Experiments on multicolor fluorescence microscopy of astrocytes on curved fiber scaffold: We apply CausWEI to single-shot 3D reconstructions of rat astrocytes grown on a curved scaffold made of electrospun fibers. The cells’ nuclei and cytoskeleton were fluorescently labelled with DAPI (4′,6-diamidino-2-phenylindole) and GFAP (Glial fibrillary acidic protein) respectively. In the in-situ microscopy of cells within tissues, it is seldom possible that all the fluorescent molecules in the field-of-view are located within the depth-of-field of a high NA objective. The electrospun fiber scaffold mimics the size scale and architecture of the natural extracellular matrix that is present in tissues. In-addition, the curvature of the scaffold replicates the natural tissue curvature, a challenging imaging condition encountered in the cortex-wide imaging of rat brain. For imaging of non-fluorescent electrospun fibers, we used scanning electron microscopy (Fig. 3a). For comparison and obtaining baseline for fluorescently labelled cells, we scanned this scaffold sample with a conventional wide-field fluorescence microscopy. The fluorescent markers in the field-of-view shown in the Fig. 3 are located in a depth range of 30 µm and hence this imaging experiment required 30 depth scans at a depth step-size of 1µm. Three of these depth-scans corresponding to nuclei and cytoskeleton markers are shown in Fig. 3. The focused regions are marked with a red arrow. Next, we captured the 3D information from the same sample and field-of-view in a single snapshot as caustic pattern with CausWEI method as shown in Figs. 3c and 3g. A reconstruction of 24 different depth-slices from this caustic snapshot was obtained and the reconstructions are shown in Supplementary Video Files 3 and 4. 3 out of these 24 depth slices of fluorescently labeled astrocyte nuclei and cytoskeleton are shown in Fig. 3.