The Materials that has been used in this experiment are as follow.
1. X-ray Imaging Detector
o sensor S10811,
o CCD detector Hamamatsu,
o Signal processing unit C9266-03.
2. X-ray Source:
o X-ray tube with adjustable voltage and current settings.
o Tube voltage (25 KV)
o Tube current (25mA)
3. Test Patterns and Phantoms:
o Bar pattern phantom for CTF measurement.
o Slit phantom for LSF measurement.
o Sharp edge phantom for ESF measurement.
3. Computer with image acquisition software.
4. Slit camera:
o Acquire an image without X-ray -Dark current image,
o Acquire an image without sample -Flat field image,
o Position the bar pattern close to the detector and make sure it is well aligned with the pixel matrix,
o Acquire an image.
5. Additional Equipment:
o For precision stages in positioning phantoms and detectors.
o The collimators, Electrometer Keithley used for controlling X-ray beam shape and size (IEC, 2015).
7. Experimental Setup
o The X-ray imaging detector was positioned at a fixed distance from the X-ray source.
o Test patterns and phantoms were placed between the X-ray source and the detector.
o Collimators were used to shape the X-ray beam as required for different measurements.
Experimental conditions
This experiment was carried out in the physics laboratory of the Physics Department of the University of Trieste. The X-ray tube, which is the main device of the experiment, is located in a chamber. The chamber was not vacuumed, so there may be changes in the temperature inside the space and the general environment over time for the test. Still, this change, in this case, we can assume that it had a negligible effect on test.
According to Williams, et al., (2007), several safety measures must be considered for the X-ray tube to operate. First of all, all the windows of the chamber must be closed. Before switching on the X-ray tube, the "kV" and "mA" knobs on the external panel should be set to 0 (zero). Turning on the POWER on the external panel only after closing the chamber is possible.
After switching on the X-ray tube, we can enter the necessary parameters in the TUBE VOLTAGE/CURRENT field of the external panel (25 kV, 40 mA). It is mandatory to close the chamber windows to switch on the X-ray tube, and then the red light inside the chamber lights up which means the tube is on.
Before the acquisition, we can check whether the parameters have reached the desired value in the TUBE VOLTAGE/CURRENT field of the external panel, and then start the process. There is a SHUTTER button on the external panel to open the X-ray tube, and after we open it, the light falls on the detector and we can record the data through the Hamamatsu C9266DCamAPL software.
ORTEC 871 Timer and Counter is used to open and close the SHUTTER button for the required time interval to maintain the accuracy of the acquisition process and uniformity in the repeatability of the experiments. The acquisition process was set to 5 seconds, and the ORTEC 871 Timer and Counter should be set to 1 second. During the exposure time of the detector, we triggered the shutter so the X-ray beam was on for 1 second.
When installing and changing samples (bar pattern, line pattern, and edge pattern) during the experiment, it is necessary to turn off the X-ray tube, and after changing the sample, restart it. It is important to highlight that to measure LSF or ESF from data in ImageJ, we need to use the measurement line of ImageJ exactly at 90 degrees to get exact values. Additionally, the correct use of the timer during the experiment, waiting until the system is fully activated after applying the current and voltage is an important factor.
i. Measurement of Contrast Transfer Function (CTF):
o The bar pattern phantom was exposed to X-rays.
o Images were acquired at various spatial frequencies.
o The contrast at each spatial frequency was measured by analyzing the intensity profiles across the bars.
o CTF was calculated as the ratio of the image contrast to the object contrast for each spatial frequency.
ii. Measurement of Modulation Transfer Function (MTF):
o The slit phantom was used to obtain the Line Spread Function (LSF).
o A narrow slit was imaged, and the intensity profile perpendicular to the slit was recorded.
o The LSF was obtained by plotting the intensity profile.
o The MTF was derived by taking the Fourier transform of the LSF.
iii. Measurement of Line Spread Function (LSF):
o The slit phantom was scanned across the detector using a precision stage.
o Multiple images were acquired to improve signal-to-noise ratio.
o The LSF was determined by analyzing the spread of the line profile.
iv. Measurement of Edge Spread Function (ESF):
o The sharp edge phantom was imaged to obtain the ESF.
o The edge response was recorded by capturing the intensity profile perpendicular to the edge.
o The ESF was plotted and differentiated to obtain the LSF.
v. Data Analysis:
o Images were processed using image analysis software.
o Profiles for CTF, MTF, LSF, and ESF were extracted from the images.
o Fourier transforms were computed where necessary to derive MTF from LSF.
o Statistical analysis was performed to ensure the accuracy and repeatability of the measurements.
vi. Validation and Calibration:
o The experimental setup was calibrated using standard procedures.
o Repeated measurements were taken to validate the consistency of the results.
o Calibration phantoms were used to ensure the accuracy of the spatial response functions.
This systematic approach ensured a comprehensive characterization of the spatial response of the X-ray imaging detector, providing valuable insights into its performance metrics.