SERIAL NUMBER.
DOP
|
AUTHOR AND TYPE OF ARTICLE
|
TIMING OF TRACHEOSTOMY AND PLACE
|
PREPARATION
|
OPERATIVE STEPS
|
post tracheostomy care
|
1.
12-05-2020
|
Menegozzo et al3
Technical note.
|
· after 14 days of intubation.
· isolated room with lower pressure than the corridor.
· If unavailable, use a room with closed doors and no laminar flow or operating room.
|
· wear a hat, foot protection, PFF3 or N95 mask and a face shield OR goggles.
· Apron and gloves (sterilized).
· additional surgical mask (in front of the N95 or PFF3) can be used.
|
· avoid electrocautery - risk of smoke formation.
· adequate paralysis.
· Stop ventilation, advance the occluded orotracheal tube and inflate the cuff below the site proposed for the tracheostomy.
· Restore ventilation, confirm adequate pre-oxygenation and then stop ventilation again, ensuring time for complete expiration, and occlude the tube.
· Create the tracheal window, minding the cuff, deflate ET and pull the tube until the tip is proximal to the tracheal window.
· After insertion of the cannula, immediately inflate the cannula cuff with the syringe already attached to the cuff channel.
· Remove the introducer and attach connection with the antiviral filter to the cannula and the circuit.
· Restore ventilation and confirm positioning with capnography (avoid using the stethoscope).
|
· use a shower for body cleaning after tracheostomy.
· Avoid humidified oxygen, use only filters.
· suctioning always in a closed circuit.
· Cuff always inflated.
· Avoid dressing changes unless there are clear signs of infection.
· Avoid switching the cannula before 7 to 10 days, giving preference to a period of lower viral load.
· Use full PPE for exchanges.
· Deflate the cuff only with the flow off and inflate the cuff of the new cannula immediately after insertion.
· Use of cannulas without fenestrations until the COVID exam is negative.
|
2.
04-Apr-20
|
Jacob et al4
letter TO EDITOR
|
· Negative pressure theatre or isolated room
· If not available consider a normal theatre with closed doors during the procedure.
· Consider turning off laminar flow [ if present].
|
· surgeons must wear full PPE.
· Consider additional protection – powered air purifying respirator-PAPR.
|
· confirm paralysis.
· Pre oxygenation with PEEP then stop ventilation and turn off flows.
· Allow time for passive expiration with open APL valve.
· consider clamping of ET then deflate cuff and advance beyond proposed tracheal window under direct vision. ensure window is sufficient size to allow easy insertion of tracheostomy tube without injury to cuff.
· insert cuffed, non-fenestrated tracheal tube.
· immediately inflate cuff and replace introducer with non-fenestrated inner tube and HME.
· establish oxygenation with PEEP again.
· confirm position with end tidal CO2 only to avoid contamination of stethoscope by auscultation.
|
· humified oxygen to be avoided, only HME.
· use only in line closed suction circuit all times.
· Periodic check cuff pressure.
· cuff should not deflated.
· do not change dressings unless frank sign of infection.
· delay first tube change at least 7-10 days.
· follow same sequence of pause in ventilation with flows off before deflating cuff and inserting new tube with immediately re-inflation of cuff and reconnection of circuit.
· cuffed non fenestrated tube to be used until patient is covid-negative.
· subsequent planned tube change at least 30 days interval.
|
3.
|
Souza Lima et al5
technical note
|
|
· Reduce team members to just the essential required professionals.
· Use the FFP3 / N95 mask, integrated visor/ full face shield / disposable sterile gowns resistant to liquids / fluids (waterproof).
· Consider using “two gloves”.
· Use a conventional tracheostome, avoiding the fenestrated models.
|
· initial advance of the endotracheal tube before the TCT window is performed to reduce aerosol elimination; • If possible, stop ventilation while the tracheal incision is being performed, and check that the cuff is already inflated before restarting ventilation; • Ventilation should be stopped before the insertion of the tracheostome, and perform a fast and accurate placement of the tracheostome with immediate inflation of the cuff; • Confirm the adequate placement, preferentially, with capnography, if available. • An HME type filter (Heat and moisture exchanger) must be placed on the tracheostome to reduce the exposure of the virus, in case of disconnection.
|
· Avoid changing the tracheostome until the patient has a COVID-19 negative.
· Only closed-circuit suction should be used.
|
4.
25-Apr-20
|
Bottia et al6
case report
|
· ICU room with negative pressure.
|
· sterile gown, cap, shoe covers, double gloves, N95 mask, goggles and face mask.
|
· Paralyse the patient.
· Establish adequate preoxygenation,
· Avoid electrocautery.
· Stop mechanical ventilation before tracheotomy.
· Deflate the cuff of the endotracheal tube.
· Push the tube forward 3 cm.
· Perform tracheotomy with cold knife.
· No tracheal or wound suctioning of blood or secretions should be attempted if not needed, to avoid aerosol generation.
· Pull the endotracheal tube just above the tracheal incision under direct vision.
· Insert the tracheal cannula, correctly inflate.
· Give ventilation again.
|
· Careful removal of personal protective equipment with supervision.
· Take a shower after procedure.
|
5.
06-04-2020
|
Pichi et al7
editorial
|
· provisional operating room set up in the ICU .
· If not available, aerial-isolated room (i.e. doors and windows closed).
|
· Cap and shoe covers.
· Mask: FFP3 (Europe) or N99 (US) mask.
· Goggles or face shield.
· use of double gown is preferable.
· use of double nitrile gloves is suggested.
|
· no more than two expert surgeons per procedure should be involved. a single anesthesist would be sufficie for mechanical ventilation and tube removal after tracheal incision.
· non-fenestrated cuffed tracheostomy tube should be used.
· deep neuromuscular blockade in order to avoid swallowing and cough reflexes.
· Once the anterior wall of the trachea is exposed, the anesthetist: – reduces the oxygen-percentage of the inflated air to 21%; – pushes the tube as caudally as possible, so as to avoid cuff breach, – hyper-inflates the tube cuff to ensure lower airway isolation.
· ventilation is stopped entirely, the tracheal tube is lifted, without extubating the patient, until its bottom side passes the tracheal window, then cannula is inserted in the tracheal lumen.
· Promptly, the cannula cuff is inflate at the appropriate pressure level and heat and moisture exchanger is positioned: only then the ventilation is resumed by close airway circuit attachment.
· Tracheal cannula should be considered correctly positioned until CO2 value is displayed, avoiding stethoscope contamination by thoracic auscultation.
|
· No dressing change should be performed unless evidence of local infection and cannula cuff should be checked regularly and not be deflated unsafely.
· The cannula should be held during any passive movement of the patient to avoid air leakage from the stoma.
· Cannula change can be planned 7–10 days later using the same standards (PPE utilization and airflow interruption). cannula change can be delayed 30 days after.
|
6.
04-Apr-20
|
Kligerman et al8
special issue
|
|
· PPE- N95 mask, face shield, gown, and gloves.
|
· Closed circuit ventilation.
· Cuffed tracheostomy tubes should be used to decrease leaks in the circuit.
· Use of in-line suction, even for patients who may not require chronic closed-circuit ventilation may also decrease risk of aerosolization and droplets.
· Nebulizer treatments should be avoided or used with caution as well to minimize risk of aerosolization.
|
· Patients not on a closed ventilation circuit should wear a surgical mask over their stoma if tolerated as this may decrease spread of droplets from leakage around the stoma and/or HME.
· tube change should be postponed and manipulation of the tracheostomy site should be minimized.
· all nonurgent clinic visits should be postponed and/or converted to telehealth visits when possible.
|
7.
31-Mar-20
|
Kowalski et al9
S P E C I A L I S S U E
|
|
· wearing PPE such as N95 or FFP2 mask or PAPR, gown, cap, eye protection, and gloves to decreasing the risk of aerosolization.
|
· stop ventilation while the tracheostomy window is being performed, and only resume ventilation when the cuff of the tracheostomy tube has been inflated.
|
|
8.
31-Mar-20
|
Vargas et al10
S P E C I A L I S S U E
|
|
· full protective wearing including N95 respirator, gown, cap, eye protection, and gloves.
|
· To avoid the aerosol, push down the endotracheal tube beyond the site chosen for the tracheal stoma at the beginning of the procedure.
· The endotracheal tube should reach the tracheal carina so the cuff is surely distal to the tracheostomy site.
|
|
9.
08-Apr-20
|
Xiao et al11
LETTER TO THE EDITOR
|
|
|
· procedures should be performed under general anesthesia, with deprivation of spontaneous respiration and application of muscle relaxants regardless of whether patients had spontaneous breathing or not, to restrain the cough reflex caused by tracheal stimulation.
· after the cervical trachea is exposed and immediately before an incision is made in the trachea, the endotracheal tube (ETT) is inserted deeper, positioned with the tip close to carina of the trachea .
· This step would prevent the ETT cuff leak due to an accidental damage to the cuff when making the tracheal opening.
· when the opening is complete, brief interruption of the ventilator is essential. Then the ETT is pulled out, and subsequently the tracheostomy tube quickly inserted into the opening almost simultaneously, the tracheostomy tube cuff is inflated and the tube rapidly connected to the ventilator with immediate resumption of the ventilator .
· Suspension of ventilation support was usually not more than 15 seconds, with satisfactory oxygen saturation.
|
|
10.
08-Apr-20
|
Miles DDS et al12
S P E C I A L I S S U E
|
· 21 days after the onset of symptoms if feasible.
· Tracheostomy should not be delayed regardless of SARS-CoV-2 status in life-saving situations or in situations in which the tracheostomy would significantly improve the prognosis of the patient.
· Ideally, the procedure should be performed at bedside in the intensive care unit in a negative pressure room or using a portable high efficiency particulate air (HEPA) filtration system to avoid patient transportation and contamination of other areas in the medical center.
· If it is necessary to perform the procedure in the operating room (OR), a specific OR cluster should be designated to avoid contamination of additional OR resources for noninfected patients.
|
· waterproof cap, goggles with an antimist screen, impermeable operating room surgeon's gown and gloves and a transparent plastic facial shield worn outside the goggles and N95.
· The minimum number of health care workers required to perform the procedure should be present to prevent unnecessary exposures.
· Consideration for power air-purifying respirator (PAPR) systems for personnel performing tracheostomy should be entertained.
|
· paralysis to prevent coughing.
· glycopyrrolate to reduce secretions.
· preoxygenation and cessation of ventilation during the tracheostomy procedure.
· utilization of closed suctioning systems.
· avoiding monopolar electrocautery, or harmonic technology, and using cold instrumentation when feasible.
· minimizing suctioning during the procedure. ensuring the
· cuff is inflated prior to resuming ventilation, the circuit should be close type.
|
· Securing circuits properly and avoiding unnecessary humidification systems may reduce the risk of unexpected circuit disconnection and aerosolization leading to exposure.
· The circuit should remain closed as much as possible, and closed-line suctioning should be used.
· Heat moister exchangers with viral filters and HEPA filtration should be used when possible.
· Tracheostomy tube changes should be avoided and should only be performed in cases of cuff failure or emergent situations.
|
11.
09-Apr-20
|
David et al13
S P E C I A L I S S U E
|
· 21 day postintubation.
· Ventilator parameters to qualify for safe tracheostomy placement include positive endexpiratory pressure (PEEP) < 12 and fraction of inspired oxygen (FiO2) < 0.60.
· ICU preferred or negative pressure room.
|
· N95 mask or PAPR, head covering, eye protection, gown, and two pairs of gloves.
|
· Limit number of providers in the room during the procedure.
· Use of paralysis to prevent coughing.
· holding ventilation when the ETT cuff is deflated and when the trachea is opened.
· Avoid use of laryngotracheal topical anesthesia.
· Holding ventilation during tracheotomy until cuff inflated and circuit reconnected.
|
· Use of closed, inline suctioning.
· Closed circuit with HEPA filter if on mechanical ventilatory support.
· HME when off ventilatory support.
· Delaying the first tracheostomy change to 1 month or after deisolation occurs for COVID-19 positive patients.
|
12.
14-Apr-20
|
Skoog et al14
S P E C I A L I S S U E
|
· >30 days from diagnosis.
· negative pressure room. Procedures should be completed in the ICU at bedside to avoid risk of exposure during transport.
|
· minimal PPE worn by staff should include N95, mask with shield, surgical gown, double gloves. PAPR.
· HEPA filter “air scrubber” should be placed in the room.
|
· Patients were completely paralyzed to minimize air movement and coughing and thus viral dissemination via aerosolization.
· Just prior to airway entry, the patients were pre-oxygenated, ventilation was held, and the cuff on the endotracheal tube was dropped to minimize air movement over the respiratory mucosa.
· While the patient was apneic, the tracheotomy incision was performed.
· Open suctioning of the trachea was avoided.
· Instead, a closed suctioning system with a viral filter was used.
· ventilation only with cuff inflation. avoiding suctioning once the trachea is incised due to the risk of aerosolization of high viral load secretions.
· minimizing cautery due to concerns of aerosolization of viral particles in the smoke plume.
|
· closed suction system with viral filter.
· Keep cuff inflated.
· Delay first tracheotomy tube change to 3-4 weeks, if possible avoid changing tracheotomy tube until after COVID has passed.
|
13.
16-Apr-20
|
Broderick et al15
|
· dedicated “Covid Theatre”, operates under negative pressure. Reverse laminar flow in the perimeter around the operating table is estimated to exchange 90% of air (removing the generated aerosol) in 6 minutes.
|
· PPE- FFP3 face mask, surgical hood, goggles or visor and double gloves.
· consider deep suctioning of the chest and oral cavity using the closed suctioning circuit prior to transfer to minimises the amount of secretions at the time of opening the trachea.
· Security staff has to close the corridor temporarily during transfers, and the anaesthetic team needs to put on PPE prior to entering ICU .
· The patient is transferred along the back corridor of the theatres and not through the main theatre entrance.
· viral filter should remain on the endotracheal tube.
|
· The patient should be fully paralysed.
· suction the ET tube again, including subglottic port .
· confirm adequate muscle relaxation.
· keep a closed circuit until the tracheostomy tube is inserted in the trachea.
· Prior to tracheal window, stop the ventilator and deflate the balloon/cuff. Afterwards, the surgeon makes the window in the trachea. Minimal suctioning is used.
· Once the window is achieved the ET tube is advanced further (past the window) and the balloon/cuff is re-inflated (over-inflated),thus establishing a closed circuit.
· At this stage the patient is ventilated and the tracheal window lies above the level of the ET cuff allowing ample time for the surgeon to check haemostasis and insert a rescue-suture.
· The ventilator is then stopped again, the cuff is deflated and the ET tube is withdrawn slowly to allow the insertion of the tracheostomy and inner tube.
· The tracheostomy cuff is inflated and the circuit is connected. The capnography CO2 trace is confirmed. The ET tube is simultaneously clamped while the tracheostomy is being placed.
|
A closed suctioning is used.
|
|
Schultza et al16
consesus
|
|
· limitation of the number of caregivers present in the operating room.
· full face shield/visor or airtight protective glasses. • FFP2 (N95) or FFP3 mask; • headlight covered by a head cap; •an impermeable protective apron or an overcoat that must be worn under the surgical gown as it is not sterile. an apron or a gown, head protection with a hood cap rather than with a simple cap in order to better prevent any skin exposure,
|
· minimize the use of electrocoagulation which can generate aerosolization of the virus when the trachea is open; • when possible, use a sterile transparent interface between the patient and the surgeon, in order to limit the risk of contamination; • if possible: carry out a drug assisted neuromuscular block to reduce any risk of coughing when opening the trachea; • stop ventilation just before the trachea is incised; Tracheostomy under local anesthesia is not recommended. However, if it is necessary, it is recommended to inject 5 cc of Lidocaine 5% intratracheally through the tracheal wall, before the incision of the trachea is performed in order to reduce the cough reflex.
|
· surgical hand scrub and/or friction with hydro-alcoholic solution before and after each treatment.
· For tracheostomy change, abundant spraying of 5% lidocaine into the tracheostomy tube, followed by an aspiration a few minutes later, is useful.
· If the patient is ventilated on the tracheostomy cannula, the anesthesiologist is asked to sedate the patient and perform a neuromuscular block to reduce any risk of coughing during the change of the cannula. All disposable material that has been in contact with the cannula or trachea (filters, suction probes) during the post-tracheostomy care must be eliminated through the infectious waste circuit. It is possible to use a room without air treatment provided that. the bedroom door is kept closed; the patient’s room is regularly ventilated; the air pressure in the room is maintained at zero. the cannula should ideally be connected to an HME filter and covered by a surgical mask.
|
14.
April 6, 2020
|
Foster et al17
novel approach
|
· negative pressure operating room.
|
· PPE-boot covers, sterile gown, gloves, surgical mask, powered air-purifying respirators(PAPRs).
· The Ecolab Scope Pillow Warmer Drape is a clear plastic material that is stretched over the retractor arms, forming a barrier between the operative field and the surgeon, while still allowing for good visualization of the operative field. The drape is then secured with snaps to the self retaining retractor to maintain the tightness of the drape; this will improve visibility. Buffalo Filter smoke evacuator tubing is connected to 2 heat moisture exchange (HME) filters and placed under the drape to provide further air filtration The operator and assistant will proceed with hands underneath the drape.
|
|
|
15.
Jul 2, 2020
|
Youn et al18
case report
|
· at bedside in a negatively pressured ICU room to minimize the risks of transmission during transfer and worsening of the patient's condition.
|
· (PPE) consisting of a Level C powered air-purifying respirator (PAPR) with an aseptic waterproof surgical gown and gloves.
|
· Avoid bipolar electrocoagulator to prevent droplet aerosolization.
· injection of sedatives and neuromuscular blockers needed.
· Prior to opening of the second tracheal ring, the FiO2 was lowered to 0.4 and the balloon was temporarily overinflated.
· The balloon was below the level of the tracheal opening.
· No droplets escaped the opening due to the repositioned and overinflated balloon. The ET tube was withdrawn until the tip of the tube reached above the tracheal opening.
· During withdrawal, to prevent the spread of droplets, pausing the ventilator was considered before balloon deflation. Immediately after ET tube tip repositioning, a hole was made on a transparent film dressing with scissors and a tracheostomy tube was immediately inserted.
· After a quick check of tracheostomy tube patency with a rubber catheter, the ventilator was connected, the tidal volume was checked, and the existing ET tube was removed.
|
|