3.1 Quantification steps
3.1.1 3D OS-EM optimization
The effect of the number of 3D OS-EM updates on the percentage recovery is illustrated in Fig. 2. The percentage recovery values were calculated from the reconstructed 177Lu SPECT images of three different sized spheres simulated in a cylindrical phantom, as well as the left and right kidneys of the patient phantom. The recovery values in Fig. 2 (a) and (b) were derived from simulations when the object-to-background ratios of 13:1 and 6:1 were used. The total counts obtained for each VOI were normalized to the maximum counts of the 3D OS-EM updates and expressed as a percentage recovery.
As seen in Fig. 2, fewer updates were required for the larger objects to reach convergence (count recovery > 90%). Convergence was reached for the right and the left kidney as well as the 65 ml sphere after only 24 updates. In contrast, the 14 ml and the 4.2 ml spheres achieved 90% recovery only after 48 and 84 updates, respectively. These observations applied to both concentration ratios in Fig. 2 (a) and (b).
3.1.2 Calibration factors
The results for the sphere-CF and the cylinder-CF were 13.9 cps/MBq and 16.3 cps/MBq, respectively. The sphere-CF underestimated the cylinder-CF by 16.3%. The results demonstrated that the VOI drawn based on the physical dimensions of the sphere excluded some of the counts for the sphere-CF due to spill-out. As a result, a lower CF value was obtained for the sphere-CF in comparison to that of the cylinder-CF.
3.1.3 Recovery coefficients
Figure 3 shows the characteristic non-linear curves fitted with R2 values of 0.99, generated using the sphere-CF and cylinder-CF, when RC values were plotted for the different sphere sizes, allowing for interpolation between the sphere sizes. The curves provided the fraction of 177Lu activity concentration recovered from the reconstructed images for the given sphere sizes, and in so doing allowed for the true concentration to be calculated, thus compensating for PVE. Although the simulated concentrations (Ctrue) were the same for all VOIs, the smaller spheres had lower calculated concentration (Ccal) values, demonstrating the PVE. It is evident from these curves that RC is strongly dependent on the size of the spheres.
The sphere-RC and cylinder-RC for the spheres and kidneys are summarized in Table 1. The RC values ranged from 0.77 to 1.00 as the size of the sphere increased from 16 ml to 65 ml. The kidney RC values were 1.18 and 0.99 for the sphere-RC and cylinder-RC, respectively. Table 1 demonstrates that there is a constant difference between the sphere-RC and the cylinder-RC values for all the simulated objects. The average sphere-RC value overestimated the average cylinder-RC value by 16.3 ± 0.04%, which may be attributed to the CF difference reported in Table 1.
Table 1 The recovery coefficient (RC) values for the sphere and kidney volumes generated for the sphere and cylinder calibration factors (CFs)
Objects
|
Sphere-RC
|
Cylinder-RC
|
% difference
|
Kidneys
|
|
|
|
L-K: 159 ml
|
1.18
|
0.99
|
16.3
|
R-K: 169 ml
|
1.18
|
0.99
|
16.3
|
Spheres
|
|
|
|
16 ml
|
0.92
|
0.77
|
16.3
|
24 ml
|
0.96
|
0.81
|
16.3
|
34 ml
|
0.98
|
0.82
|
16.3
|
65 ml
|
1.00
|
0.84
|
16.4
|
|
|
Average
|
16.3 ± 0.04
|
L-K, left kidney; R-K, right kidney; RC, recovery coefficient
3.2 Evaluation of quantification accuracy
Cylindrical phantom
Table 2 shows the quantification error results for the spheres simulated in the cylindrical phantom with and without PVC. The results were calculated using the sphere-CF and cylinder-CF without PVC, while the combinations of sphere-CF-RC and cylinder-CF-RC were applied to compensate for the PVE.
Table 2 Quantification error results for spheres simulated in a cylindrical phantom with and without partial volume corrections (PVCs)
|
Quantification error (%)
|
Object
|
Without PVC
|
With PVC
|
Without PVC
|
With PVC
|
Sphere
|
Sphere-CF
|
Sphere-CF-RC
|
Cylinder-CF
|
Cylinder-CF-RC
|
16 ml
|
-4.56
|
3.32
|
-20.08
|
3.28
|
24 ml
|
-3.40
|
0.19
|
-19.11
|
0.19
|
34 ml
|
-2.05
|
-0.43
|
-17.98
|
-0.43
|
65 ml
|
1.60
|
0.83
|
-14.93
|
0.86
|
Average
|
-2.10 ± 2.67
|
0.98 ± 1.64
|
-18.03 ± 2.23
|
0.98 ± 1.62
|
CF, calibration factor; PVC, partial volume correction; RC, recovery coefficient
As seen in Table 2, the trend for the quantification error without PVC was as expected. The smallest sphere showed the largest quantification error, demonstrating the influence of PVEs, which was less important with increasing sphere size. The average quantification error obtained with the sphere-CF without PVC was -2.10 ± 2.67% in comparison to that obtained with the cylinder-CF by a value of -18.03 ± 2.23%. The use of the cylinder-CF without PVC underestimated the true activity evidently for all the sphere sizes. PVC improved the average quantification error dramatically from -18.03 ± 2.23% to 0.98 ± 1.62% for the cylinder results. The precision (standard deviation) of the average quantification error improved slightly with PVC for both the sphere and cylinder data. A slight overestimation (≤ 3.32%) of the activity values was observed for both phantoms for the smallest spheres when PVC was applied. The use of the sphere-CF-RC and cylinder-CF-RC resulted in comparable average quantification errors of 0.98 ± 1.64% and 0.98 ± 1.62%, respectively.
Torso phantom
The quantitative error results obtained for the spheres simulated in the torso phantom are demonstrated in Table 3. Similar to the cylindrical phantom, the results were analyzed using the two sphere-CF and cylinder-CF without PVC, and the corresponding sphere-RC and cylinder-RC were applied to correct for partial volume.
Table 3 Quantification error results for spheres simulated in the torso phantom with and without partial volume corrections (PVCs)
|
Quantification error (%)
|
Object
|
Without PVC
|
With PVC
|
Without PVC
|
With PVC
|
Sphere
|
Sphere-CF
|
Sphere-CF-RC
|
Cylinder-CF
|
Cylinder-CF-RC
|
16 ml
|
-4.92
|
2.92
|
-20.39
|
2.99
|
24 ml
|
-5.54
|
-2.03
|
-20.90
|
-2.03
|
34 ml
|
-4.84
|
-3.26
|
-20.32
|
-3.26
|
65 ml
|
-2.70
|
-3.44
|
-18.53
|
-3.41
|
Average
|
-4.50 ± 1.24
|
-1.45 ± 2.98
|
-20.04 ± 1.04
|
-1.43 ± 3.01
|
CF, calibration factor; PVC, partial volume correction; RC, recovery coefficient
Without PVC, the quantification error showed an average underestimation of 4.50 ± 1.24% for the sphere-CF. Similar to the sphere quantification results in the cylindrical phantom; the cylinder-CF considerably underestimated the quantified concentration, with an absolute average of 20.04 ± 1.04% for all sphere sizes.
It can be seen from the results in Table 2 and Table 3 that the calculated concentration underestimated the true concentration to the same extent between the two phantoms, with the worst quantification results obtained when applying the cylinder-CF with no PVC. Partial volume correction for both the sphere and cylinder slightly overestimated the concentration of the two smallest spheres (≤ 2.99%). Analogous to the sphere quantification in the cylindrical phantom, the application of the sphere-CF-RC and cylinder-CF-RC resulted in comparable average quantification errors of -1.45 ± 2.98% and -1.43 ± 3.01%, respectively.
Patient phantom
Figure 4 shows the patient phantom reconstructed SPECT data with its associated CT data illustrating the sphere and the kidney objects (indicated by the arrows) used for quantification analysis. Table 4 compares the quantification errors obtained for the spherical and kidney objects simulated in the patient phantom calculated using the sphere and the cylinder data.
As seen from Table 4, the sphere quantification errors showed similar trends to those obtained in the cylindrical and torso phantoms. A small average quantification error of -0.93 ± 2.97% was obtained when applying the sphere-CF without PVC in the patient phantom. Partial volume correction using the sphere-CF-RC altered the average quantification error to 2.18 ± 0.91%. These results were comparable to those obtained in the torso (-1.45 ± 2.98%) and the cylindrical phantom (0.98 ± 1.64%). The cylinder-CF underestimated the absolute average quantified concentration of the spheres (17.07 ± 2.48%) similar to the torso (20.04 ± 1.04%) and the cylindrical phantom (18.03 ± 2.23%). Partial volume correction improved the cylinder-CF-RC to an average of 2.20 ± 0.94% similar to the torso (-1.43 ± 3.01%) and cylindrical phantom (0.98 ± 1.62%). When all corrections were applied, the sphere-CF-RC (2.18 ± 0.91%) was comparable to the cylinder-CF-RC (2.20 ± 0.94%) data, validating findings of the cylindrical and torso phantoms.
Table 4 Quantification error results for the spherical and kidney objects simulated in the patient phantom calculated with and without partial volume corrections (PVCs)
|
Quantification error (%)
|
Objects
|
Without PVC
|
With PVC
|
Without PVC
|
With PVC
|
Spheres
|
Sphere-CF
|
Sphere-CF-RC
|
Cylinder-CF
|
Cylinder-CF-RC
|
16 ml
|
-4.42
|
3.47
|
-19.97
|
3.53
|
24 ml
|
-1.70
|
1.95
|
-17.69
|
1.94
|
34 ml
|
-0.34
|
1.32
|
-16.65
|
1.32
|
65 ml
|
2.73
|
1.96
|
-13.98
|
1.99
|
Average
|
-0.93 ± 2.97
|
2.18 ± 0.91
|
-17.07 ± 2.48
|
2.20 ± 0.94
|
Kidneys
|
|
|
|
|
L-K: 159 ml
|
18.98
|
1.01
|
-0.37
|
0.94
|
R-K: 169 ml
|
22.34
|
3.50
|
2.44
|
3.45
|
Average
|
20.66 ± 2.38
|
2.26 ± 1.76
|
1.04 ± 1.99
|
2.20 ± 1.77
|
CF, calibration factor; L-K, left kidney; PVC, partial volume correction; R-K, right kidney; RC, recovery coefficient
The kidney quantification showed an average overestimation of 20.66 ± 2.38% for the sphere-CF without PVC, in comparison to the better results (1.04 ± 1.99%) obtained for the cylinder-RC. This was contrary to the sphere quantification results where a considerable underestimation of the quantification error was obtained using the cylinder-CF without PVC. The kidney quantification error improved to 2.26 ± 1.76% with the sphere-CF-RC corrections. With PVC the sphere-CF-RC and cylinder-CF-RC yielded comparable average quantification results of 2.26 ± 1.76% and 2.20 ± 1.77%, respectively; similar to the findings of the sphere quantification results. The results suggest that PVC did not play a major role in the kidney quantification error using the cylinder data. This may be attributed to the fact that the cylinder-RC value applied to the kidneys was close to 1.0.